Real-time statistics

Reliable actuator

Minimal maintenance

Model MDV Heart Valve Durability Tester

A touchscreen HMI (human-machine interface) allows the user to enter parameters to control the test. Most parameters are electroncialy settable and can be changed while the test is running.

Differential-pressure statistics (relevant to regulatory guidance) are calculated and displayed immediately. Data can be exported at user specified intervals and sent to PC-based data acquisition software.

Auto-tuning mode automatically adjusts pump amplitude to achieve set heart valve differential pressure.

Calibration and checking of pressure, and temperature is very quick and easy. Specific screens for checking or calibrating walk the user through the process. The intuitive process reduces the likelihood of calibration mistakes.

The MDV machine requires no maintenance. The fluid portion of the machine can be cleaned with IPA or water.

Required Installation Connections – Model MDV
Description Physical Specs Functional Specs
Electric Power Powers the control system, HMI, actuator, heaters, etc. C13 Coupler (IEC 60320) 100-125 VAC 3A 50-60 Hz

220-240 VAC 3A 50-60Hz

Compressed Air Provids air for systemic pressure system, etc.
(see Knowledge Base: Compressed Air &
High Pressure Gas Connections
)
Instant Tube Fitting Accepts ¼-inch OD Plastic Tube 1.4 to 8.6 bar (20 to 125 psi)

Model MDV Heart Valve Durability Tester

Easy-to-use durability tester makes it quick to set up test and collect data. Custom valve holder can accommodate variety of heart valve designs.

Test Frequency Temperature Control Accuracy
5 – 25 Hz +/- 1 deg C