Real-time statistics
Reliable actuator
Minimal maintenance
Model MDV Heart Valve Durability Tester

AÂ touchscreen HMIÂ (human-machine interface) allows the user to enter parameters to control the test. Most parameters are electroncialy settable and can be changed while the test is running.
Differential-pressure statistics (relevant to regulatory guidance) are calculated and displayed immediately. Data can be exported at user specified intervals and sent to PC-based data acquisition software.
Auto-tuning mode automatically adjusts pump amplitude to achieve set heart valve differential pressure.
Required Installation Connections – Model MDV | ||||
---|---|---|---|---|
Description | Physical Specs | Functional Specs | ||
Electric Power | Powers the control system, HMI, actuator, heaters, etc. | C13 Coupler (IEC 60320) | 100-125 VAC 3A 50-60 Hz
220-240 VAC 3A 50-60Hz |
|
Compressed Air | Provids air for systemic pressure system, etc. (see Knowledge Base: Compressed Air & High Pressure Gas Connections) |
Instant Tube Fitting Accepts ¼-inch OD Plastic Tube | 1.4 to 8.6 bar (20 to 125 psi) |
Model MDV Heart Valve Durability Tester
Easy-to-use durability tester makes it quick to set up test and collect data. Custom valve holder can accommodate variety of heart valve designs.
Test Frequency | Temperature Control Accuracy |
---|---|
5 – 25 Hz | +/- 1 deg C |